MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE~RITE8 ULTRASOUND SYSTEM 20MM CUE LINEAR PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE~RITE8 ULTRASOUND SYSTEM 20MM CUE LINEAR PROBE; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770603
Device Problems Image Display Error/Artifact (1304); Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed - black mark in the center of the screen that impairs vision.
 
Event Description
Per biomed - black mark in the center of the screen that impairs vision.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of a black mark in the center of the screen that impairs vision was confirmed; the probe displays a black thick line in the middle.It's as if it is not scanning in the middle of the probe.The root cause of the reported failure was identified as a probe failure.A history review of serial number asduag102 showed no other similar product complaint(s) from this serial number.Evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE8 ULTRASOUND SYSTEM 20MM CUE LINEAR PROBE
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11322837
MDR Text Key233327779
Report Number3006260740-2021-00335
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138218
UDI-Public(01)00801741138218
Combination Product (y/n)N
PMA/PMN Number
K182281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770603
Device Catalogue Number9770603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2021
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/12/2021
Supplement Dates Manufacturer Received06/07/2021
Supplement Dates FDA Received06/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-