ZIMMER BIOMET, INC. TPRLC XR T1 PPS 14X148MM; PROSTHESIS HIP Back to Search Results
Catalog Number 51-105140
Device Problems Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/28/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the surgeon noticed foam inside the package from the stem pouch tearing and the stem rubbing the inside of the package.The surgery was completed with a backup device.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4;g3; h2; h3; h4; h6 visual evaluation of the returned product/provided photo(s) identified damage to the sterile packaging pouch with debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging and the porous coating.Sterility has not been compromised.Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.Device history record (dhr) was reviewed and no discrepancies were found.The root cause of the reported event can be attributed to transit damage.The sterile packaging configuration is moving to a double blister configuration if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 14X148MM
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11324323
MDR Text Key231701551
Report Number0001825034-2021-00461
Device Sequence Number1
Product Code OQG
Combination Product (y/n)N
PMA/PMN Number
K1200330
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number51-105140
Device Lot Number6222399
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/02/2021
Initial Date FDA Received02/12/2021
Supplement Dates Manufacturer Received05/10/2021
Supplement Dates FDA Received05/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-