• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET EXACT ASSY, 45 DEGREE X 14MM; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET EXACT ASSY, 45 DEGREE X 14MM; LAVAGE, JET Back to Search Results
Model Number 66800041
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2021
Event Type  malfunction  
Event Description
It was reported that during set up the versajet hand piece was found faulty.Irrigation connection had a perforation, occurring in spillage as it was connected to irrigation.There was a delay of 5 minutes and a backup was available to start with the treatment.No patient harm reported.
 
Manufacturer Narrative
The device, was not used in a procedure was not returned for evaluation with all additional information provided we have not been able to establish a relationship between the reported event or determine a root cause.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found other related failures.Probable cause may be a component failure or connection issue.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET EXACT ASSY, 45 DEGREE X 14MM
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11326044
MDR Text Key231937672
Report Number8043484-2021-00299
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00040565124681
UDI-Public00040565124681
Combination Product (y/n)N
PMA/PMN Number
K110958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66800041
Device Catalogue Number66800041
Device Lot Number50782235
Was Device Available for Evaluation? No
Date Manufacturer Received03/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-