The patient's sample was returned for testing.The customer's results were reproduced and interference testing found no interfering factors.The calibration was acceptable.Only the 3rd level of control was measured on the day of the event.The sample was found within 1 sd range.To the differences of the ft4 values generated with the different analyzers: it needs to be taken into account that assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used and, differences in reference materials/methods, and the standardization methodology used.A general reagent issue most likely can be excluded.No interference has been found within the sample.The different values are most likely based on methodological differences.
|