• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH HEX END ROD, 480MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH HEX END ROD, 480MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Catalog Number 179762480
Device Problem Migration (4003)
Patient Problem Unspecified Tissue Injury (4559)
Event Type  Injury  
Manufacturer Narrative
510k: this report is for an unknown rod/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2020, a primary fusion procedure at l1~sai was performed.Verse system was applied to l1/s, and expedium sai was used at sai.On (b)(6) 2021, they found that the rod which was deployed at l1 right had come off the screw¿s head.Additionally, the setscrew had been placed in the screwhead.During the revision procedure, the fusion was extended up to th10 without any issue.No further information is available.This report is for one (1) unknown rods.This is report 3 of 3 for (b)(4).
 
Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: updated concomitant device reported.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d6a, d6b: updated.H3, h6: the investigation could not be completed, no conclusion could be drawn at the time of filing this report.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: the dhr of product code 179762480, lot bdn2es2, was reviewed and no non-conformances were observed during the manufacturing process.The product was released on january 16, 2020.Qty.150.The dhr was electronically reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEX END ROD, 480MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11332553
MDR Text Key231997234
Report Number1526439-2021-00302
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
PMA/PMN Number
K111136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number179762480
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/16/2021
Supplement Dates Manufacturer Received02/18/2021
04/01/2021
Supplement Dates FDA Received03/16/2021
04/13/2021
Patient Sequence Number1
Treatment
5.5 EXP VERSE CAN SCR 6.0X45; 5.5 EXP VERSE CAN SCR 6.0X45; 5.5 EXP VERSE UNITIZED SET SCR; 5.5 EXP VERSE UNITIZED SET SCR; ROD, 480 MM; 5.5 EXP VERSE CAN SCR 6.0X45; 5.5 EXP VERSE UNITIZED SET SCR
Patient Outcome(s) Required Intervention;
-
-