Catalog Number 00880100100 |
Device Problem
Unstable (1667)
|
Patient Problems
Abrasion (1689); Laceration(s) (1946)
|
Event Date 10/19/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Final report will be filed upon completion of the investigation.
|
|
Event Description
|
It was reported that during surgery the power tool was not working and additional unit was used to complete procedure.When in use the hand piece skipped and took a graft too deep for use, so the skin had to be stitched back in place.There were no additional consequences to the patient.No additional information available.
|
|
Event Description
|
No additional event information available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the unit took a skin graft that was too deep and was out of calibration.The unit was calibrated and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|
|