Model Number CI-1600-05 |
Device Problems
Therapeutic or Diagnostic Output Failure (3023); Migration (4003)
|
Patient Problems
Therapeutic Response, Decreased (2271); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 02/04/2021 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced decreased performance.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts to the silicone on the top cover.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis of the electrode pocket did not reveal any corrosion in the feed-thru pocket.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|