Model Number V60
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2021.Reported: 18feb2021.
 
Event Description
The customer called into technical support (ts) reporting that the device is displaying a blower over temperature message.The device was in clinical use at the time the reported issue was discovered, however, there was no interruption in therapy or delay as it had not started yet.There was no harm to the patient or user.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed reported problem.The cause of the blower overheating error in error log was that both fan and blower filters were clogged causing the motor to overheat.The fse replaced the fan and blower filters to resolve the reported issue.The unit passed the required performance verification tests per philips standards and was put back into service.
 
Manufacturer Narrative
G4:02mar2021.B4:10mar2021.The customer reported there was no patient involvement at the time the issue was discovered.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11343373
MDR Text Key233976758
Report Number2031642-2021-00642
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2021
Initial Date FDA Received02/18/2021
Supplement Dates Manufacturer Received01/22/2021
Supplement Dates FDA Received03/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-