DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Gas/Air Leak (2946)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit and the o-ring on the rear of the pump console was found to be broken.The o-ring was replaced to fix the issue, and two new helium tanks were provided as a customer courtesy.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during an in-service training performed by a getinge clinical representative, it was observed that the brand new helium tank on the cardiosave intra-aortic balloon pump (iabp) was completely empty.When the tank was replaced with a full one, there was an audible helium leak.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, h2, h6, h10.The device history record (dhr) for this iabp was reviewed and no non-conformances related to the reported event were noted.
|
|
Event Description
|
It was reported that during an in-service training performed by a getinge clinical representative, it was observed that the brand new helium tank on the cardiosave intra-aortic balloon pump (iabp) was completely empty.When the tank was replaced with a full one, there was an audible helium leak.There was no patient involvement, and no adverse event reported.
|
|