A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed that the iabp unit was infested with saline which was indicated by white traces of salt on the walls and bottom of the card after the console had been pulled out.It was noted that the iabp unit could not be switched on even after the 230 vac mains voltage was connected.Further troubleshooting revealed a non-functional power supply monitor board.The measured value of the output voltage at the power supply terminals was 2.3 vdc.The fse disconnected the j1 connector on the power supply monitor board and observed a voltage value of 15.1 vdc measured at the power supply terminals.The fse recommended the customer replace the power supply monitor board.The fse completed pm service on the iabp unit, however, the iabp unit was not cleared for use.Subsequently, the fse returned to the customer's site at a later date and replaced the power supply monitor board.The iabp unit was observed to turn on.The fse then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.Updated fields: b4, g4, g7, h2, h6 (evaluation result codes, evaluation conclusion codes) h10, h11 corrected fields: a1, a3, b5, d1, h6 (patient codes, device codes), h10.
|
Updated fields: b4, g3, g6, h2, h4, h6 (type of investigation), (component codes), h10.
|