DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
During a preventative maintenance (pm) service, the getinge field service engineer (fse) discovered a broken fiber optic connector.To fix the issue, the fse installed a new fiber optic sensor assembly.The fse then performed calibration, all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.The initial reporter named in is a getinge employee who has different contact details from that of the event site, and has the following contact information: (b)(6).The full event site name in is (b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the fiber optic connector of the cardiosave intra-aortic balloon pump (iabp) was found broken.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the fiber optic connector of the cardiosave intra-aortic balloon pump (iabp) was found broken.There was no patient involvement, and no adverse event reported.
|
|