• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS PCA MODULE; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS PCA MODULE; PUMP, INFUSION Back to Search Results
Model Number 8120
Device Problems Break (1069); Crack (1135); Deformation Due to Compressive Stress (2889); Contamination /Decontamination Problem (2895); Device Markings/Labelling Problem (2911); Appropriate Term/Code Not Available (3191)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The affected device has been received and an evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
It was reported that the device was broken / damaged.There was no patient involvement.
 
Event Description
It was reported that the device was broken / damaged.There was no patient involvement.
 
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that they replaced front cover, rear case, left/ right handles, front/ rear doors, barrel clamp, lock label, latch module, tube drive, sleeve drive, wiper seal, flange gripper retainer, small/ large claws and left/ right iui's.Performed calibration.The failure code othe was used to track the alaris pump software version as received from the customer and the software version when device was sent back to the customer.It does not reflect a device failure or represent any risk to the patient.A review of the device history record showed the device had a manufacture date of 25jan2012.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.Based on the findings, service determined that the probable root cause of the reported issue was due to wear of the front case - damaged/ cracked (front cover bottom left corner).A review of the complaint history record in trackwise and sap was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PCA MODULE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key11355773
MDR Text Key232865080
Report Number2016493-2021-26787
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10885403812002
UDI-Public10885403812002
Combination Product (y/n)N
PMA/PMN Number
K043299
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8120
Device Catalogue Number8120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/19/2021
Supplement Dates Manufacturer Received05/26/2021
Supplement Dates FDA Received05/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2720-2020
Patient Sequence Number1
-
-