Model Number HL-90
Device Problem Failure to Calibrate (2440)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation- the device was returned for evaluation.The device was given safety/electrical testing and was confirmed the device was out of calibration.The device was successfully calibrated during service.
 
Event Description
It was reported the device could not be calibrated.Issue was found during calibration- no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11356582
MDR Text Key232832454
Report Number3012307300-2021-01406
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/21/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-