• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON UNSPECIFIED BD CONNECTA; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON UNSPECIFIED BD CONNECTA; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number UNKNOWN
Device Problems Complete Blockage (1094); Leak/Splash (1354); Free or Unrestricted Flow (2945); Infusion or Flow Problem (2964); Packaging Problem (3007)
Patient Problems Air Embolism (1697); Unspecified Infection (1930)
Event Date 01/25/2021
Event Type  Injury  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used.Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A catalog and lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.Device manufacture date: unknown.(b)(4).Investigation summary: bd was unable to perform a thorough investigation as no sample, lot, or batch number were provided.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Per bd corrective and preventive action (capa) corporate procedure, the reported issue does not represent a single significant incident that would trigger a capa.
 
Event Description
It was reported that unspecified bd¿ connecta had air in the line, flow issues, leakage, disconnected stopcock, kinked tubing, and open package.The following information was provided by the initial reporter: material no: unknown batch no: unknown.It was reported via post market survey that the clinician encountered occurrences of infection (blood stream bacteraemia, sepsis etc.) (10), air embolism (1), no flow or reduced flow of fluids (45), free or unrestricted flow (1), leakage (3), stopcock inadvertently disconnects from mating component (49), tubing disconnects from stopcock (11), tubing kinked (33), and package seal open before use (1).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD CONNECTA
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key11358318
MDR Text Key241451486
Report Number2243072-2021-00495
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2021
Initial Date FDA Received02/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-