Model Number 0998-00-3023-53
Device Problem Connection Problem (2900)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp, but was unable to reproduce the reported issue.The fse tested the unit without any issues.The fse performed full calibration, all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.(b)(6).
 
Event Description
It was reported that during a bypass surgical procedure, the cs300 intra-aortic balloon pump (iabp) alarmed "disconnect" error.It was also reported that the end user checked and secured all lines and connections, before ultimately switching to another maquet iabp unit, without any issues, to continue patient treatment.Further, the unit involved was sent to the biomedical customer awaiting evaluation.No patient harm, serious injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11359522
MDR Text Key233257070
Report Number2249723-2021-00341
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SENSATIO-PLUS 50CC SN#(B)(6)
-
-