Model Number 0684-00-0480-07 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complete initial reporter name: (b)(6).Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that during intra-aortic balloon (iab) therapy, the customer found a leak and the iab appeared to be ruptured.The iab was immediately replaced with a new one.There was no patient harm or adverse event reported.
|
|
Event Description
|
It was reported that during intra-aortic balloon (iab) therapy, the customer found a leak and the iab appeared to be ruptured.The iab was immediately replaced with a new one.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint (b)(4).
|
|
Manufacturer Narrative
|
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and traces of blood found on the exterior of the catheter.No blood was observed inside the iab catheter.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The reported event cannot be confirmed by the evaluation.A device and lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period mar-19 through feb-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #: (b)(4).
|
|