ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Skin Inflammation/ Irritation (4545); Contact Dermatitis (4546)
Event Date 02/08/2021
Event Type  Injury  
Manufacturer Narrative
At this time, product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dhrs (device history record) for the freestyle libre kit were reviewed, and the dhrs showed the freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The date the incident occurred is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported on behalf of the customer a skin reaction at the site of the freestyle libre 2 sensor, with symptom of dermatitis.The customer had contact with family doctor and was prescribed unspecified medication for treatment.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Sensor ((b)(6)) has been returned and investigated.Visual inspection was performed on the returned sensor patch and adhesive, no issues were observed.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A caller reported on behalf of the customer a skin reaction at the site of the freestyle libre 2 sensor, with symptom of dermatitis.The customer had contact with family doctor and was prescribed unspecified medication for treatment.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key11373936
MDR Text Key233445275
Report Number2954323-2021-05059
Device Sequence Number1
Product Code QLG
UDI-Device Identifier00357599800000
UDI-Public00357599800000
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2021
Device Model Number71992-01
Device Catalogue Number71992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2021
Initial Date Manufacturer Received 02/08/2021
Initial Date FDA Received02/24/2021
Supplement Dates Manufacturer Received05/04/2021
Supplement Dates FDA Received05/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient Weight75
-
-