Model Number 71173377
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Event Description
It was reported that, while assembling trays, it was noticed that one of the tips of the rdce frcps w/ pts brd was bent.A back up device was available, if needed.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.The visual inspection of the returned forceps confirms the tip is bent causing the stated failure.This device shows signs of significant wear and usage.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship between the device and the reported incident was corroborated.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RDCE FRCPS W/ PTS BRD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11376959
MDR Text Key233440712
Report Number1020279-2021-01587
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010519535
UDI-Public03596010519535
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71173377
Device Catalogue Number71173377
Device Lot Number20DTCD044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2021
Date Manufacturer Received05/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-