ANGIOMED GMBH & CO. MEDIZINTECHNIK KG FLUENCY PLUS VASCULAR STENT GRAFT Back to Search Results
Catalog Number FVM08040
Device Problems Difficult or Delayed Positioning (1157); Fracture (1260); Retraction Problem (1536)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.The catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified.(expiry date: 10/2022).
 
Event Description
It was reported that during stent placement procedure, the outer sheath of the stent allegedly could not be withdrawn after multiple attempts.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft products that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft products are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not available for evaluation.Based on the provided images, sheath fracture can be confirmed.In this case the system was flushed but the product was used off label.Based on the information available, the investigation is closed with confirmed result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential factors.Regarding preparation the instructions for use states: 'prior to loading the delivery system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.' furthermore, the instructions for use states: 'prior to stent graft deployment (.), ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy.' the product is indicated for use in the iliac and femoral arteries.H10: d4 (expiry date: 10/2022), g3.H11: h6 (device, method, result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during stent placement procedure, the stent allegedly difficult to deploy.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS VASCULAR STENT GRAFT
Type of Device
VASCULAR STENT GRAFT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key11377400
MDR Text Key245117721
Report Number9681442-2021-00065
Device Sequence Number1
Product Code PFV
Combination Product (y/n)N
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFVM08040
Device Lot NumberANDX5083
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/29/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received06/14/2021
Supplement Dates FDA Received06/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
Patient Weight55
-
-