• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Patient Involvement (2645)
Event Date 02/04/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched for a scheduled preventative maintenance (pm) service and resolved the reported issue by replacing the batteries.The fse then performed a pm, all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.The initial reporter named in block e1 is a getinge employee who has different contact details from that of the event site, and has the following contact information: (b)(6).
 
Event Description
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11377733
MDR Text Key233455349
Report Number2249723-2021-00392
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/04/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received04/08/2021
Supplement Dates FDA Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-