Model Number N/A |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched for a scheduled preventative maintenance (pm) service and resolved the reported issue by replacing the batteries.The fse then performed a pm, all functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.The initial reporter named in block e1 is a getinge employee who has different contact details from that of the event site, and has the following contact information: (b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) failed the battery runtime test.There was no patient involvement, and no adverse event reported.
|
|