• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 02/04/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the unit and observed the reported issue.The fse resolved the issue by replacing both batteries that were customer purchased parts and allowed the unit to charge for 48 hours.The unit was under observation, thereafter.Further, the fse performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.
 
Event Description
It was reported that during a routine check, the battery module of the cs100 intra-aortic balloon pump (iabp) was found to require replacement.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during a routine check, the battery module of the cs100 intra-aortic balloon pump (iabp) was found to require replacement.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11378585
MDR Text Key233574189
Report Number2249723-2021-00399
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public10607567107745
Combination Product (y/n)N
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/04/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received04/02/2021
Supplement Dates FDA Received04/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-