Model Number N/A |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the unit and observed the reported issue.The fse resolved the issue by replacing both batteries that were customer purchased parts and allowed the unit to charge for 48 hours.The unit was under observation, thereafter.Further, the fse performed all functional and safety checks to meet factory specifications.The iabp was then released to the customer for return to clinical service.
|
|
Event Description
|
It was reported that during a routine check, the battery module of the cs100 intra-aortic balloon pump (iabp) was found to require replacement.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during a routine check, the battery module of the cs100 intra-aortic balloon pump (iabp) was found to require replacement.There was no patient involvement, and no adverse event reported.
|
|