• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 02/11/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported issue.To fix the issue, the fse replaced the power supply and 9v batteries, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that prior to use, both batteries of the cs300 intra-aortic balloon pump (iabp) were defective.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, b6, b7, d11, g3, g4, g7, g8, h2, h6(component codes), h10, h11 corrected fields: d1, d4(model#), g1(contact person), h4.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key11382127
MDR Text Key233749456
Report Number2249723-2021-00437
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/11/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-