Model Number N/A |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The getinge field service engineer (fse) that encountered the issue replaced the lcd display and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) lcd monitor was blurry and hard to recognize characters.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) lcd monitor was blurry and hard to recognize characters.There was no patient involvement, and no adverse event reported.
|
|