Manufacturer Narrative
(b)(6).The getinge field service engineer (fse) that encountered the issue replaced the lcd display and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) lcd monitor was blurry and hard to recognize characters.There was no patient involvement, and no adverse event reported.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge field service engineer (fse), the cs300 intra-aortic balloon pump (iabp) lcd monitor was blurry and hard to recognize characters.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key11382476
MDR Text Key233750171
Report Number2249723-2021-00438
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-