Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Occlusion (1984)
Event Type  Injury  
Manufacturer Narrative
On completion of the investigation a follow up report will be submitted.
 
Event Description
Article received: mcelligott, h.E.(2020).Hybrid covered endovascular reconstruction of the aortic bifurcation (cerab) procedure is preferable to aortobifemoral bypass for limb-threatening aoartoiliac occlusive disease during the covid-19 crisis.Journal of vascular surgery cases and innovative techniques, 703-706.Purpose: this case report describes the management of a patient with complex aortoiliac occlusive disease by a hybrid endovascular approach in light of these constraints, with a successful outcome.Method: a case study per the article adverse events included occlusion.
 
Event Description
N/a.
 
Manufacturer Narrative
Article reviewed: mcelligott, et al.(2020).Hybrid covered endovascular reconstruction of the aortic bifurcation (cerab) procedure is preferable to aortobifemoral bypass for limb-threatening aortoiliac occlusive disease during the covid 19 crisis.J cardiovasc surg, cases and innovative techniques; 6:703-706.The subject article is a case report of 57-year-old man with acute-on-chronic transatlantic inter-society consensus ii (tasc ii) d aortoiliac occlusive disease managed by covered endovascular reconstruction of the aortic bifurcation as an alternative to open surgery in the era of covid-19.This complaint is based on information found within a article/literature review.There was no product that was available for evaluation, therefore a device evaluation could not be conducted and the complaint cannot be confirmed.The author of the article did not report any major adverse patient effects as result of this event.A device history record (dhr) review was unable to be performed as the device product part number and lot number was not provided within the article.Attempts to obtain the device lot information was conducted but unsuccessful.The hazardous situation/harm is addressed in the risk file and is operating within its risk profile.There was no evidence within the article that the device was the cause of the reported event.The complaint history review did not identify an adverse trend, therefore no escalation to capa process is required.Conclusion: although the right internal iliac was unintentionally occluded and right superficial femoral artery embolectomy was necessary because of acute embolus, however the patient did not require a critical care bed postoperatively, his rest pain resolved and he was discharged home.After review of the details provided, that clearly highlights the possible risks and complications known to occur following advanta v12tm balloon expandable covered stent implantation, one can infer the unfortunate injuries suffered by this patient are potentially multifactorial and an atrium advanta v12tm balloon expandable covered stent was not the only attributing factor.The factors likely include but are not limited to, the patient¿s (tasc ii) d aortoiliac occlusive disease, preexisting occlusion of left internal iliac artery and the large-volume right common iliac artery aneurysm thrombus.The authors do not attribute development of complications to any particular stent type.H3 other text : not available for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA V12 COVERED STENT
Type of Device
STENT, ILIAC
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer (Section G)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer Contact
lori gosselin
40 continental blvd
merrimack, NH 
MDR Report Key11382573
MDR Text Key233725603
Report Number3011175548-2021-00219
Device Sequence Number1
Product Code NIO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2021
Initial Date FDA Received02/25/2021
Supplement Dates Manufacturer Received11/16/2022
Supplement Dates FDA Received11/22/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexMale
-
-