BOSTON SCIENTIFIC CORPORATION AUTOGEN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G177 |
Device Problems
High impedance (1291); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
|
Event Description
|
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) heard beep tones and presented at the clinic.Interrogation was performed on the device and noted high right ventricular (rv) pacing impedance out of range, measuring greater than 3000 ohms.Lead testing was completed and showed normal rv impedance at 495 ohms.The device was not reprogrammed and the settings remained the same.The decision was made to continue monitoring this patient.The patient was advised to contact the clinic if beep tone are heard again.No adverse patient effects were reported.This device remains in service.The rv lead is a competitor lead.
|
|
Search Alerts/Recalls
|
|
|