Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2021
Event Type  malfunction  
Manufacturer Narrative
The country of origin is (b)(6).(b)(4).A general reagent issue can be excluded.Assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and differences in the standardization methodology used.The investigation did not identify a product problem.
 
Event Description
The initial reporter complained of questionable elecsys ft4 iii assay results for 1 patient sample on a cobas 6000 e 601 module analyzer with an unknown serial number and a cobas 8000 e602 serial number (b)(4).The customer's result on the e601 was 1.72 ng/dl and the result on the centaur was 1.34 ng/dl.The sample was provided for investigation where it was tested on an e602 on the (b)(6) 2021 and the result was 1.51 ng/dl.The customer reported the results outside the laboratory.The ft4 reagent lot number used on the e601 by the customer was asked for but not provided.The ft4 reagent lot number used on the e602 by the investigation team was 478085.The expiration date was may 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key11388422
MDR Text Key253614670
Report Number1823260-2021-00601
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot NumberASKU, 478085
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/04/2021
Initial Date FDA Received02/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age71 YR
-
-