• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII ARTICULAR INSERTER/EXTRACT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII ARTICULAR INSERTER/EXTRACT; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Model Number 71440194
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2021
Event Type  malfunction  
Event Description
It was reported that during an office inspection, gii articular insert/extract was noticed to have the tip bent.No case involved.
 
Manufacturer Narrative
Additional information: d4/ h4.Results of investigation: the associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tip of the returned device is deformed rendering the device inoperable.The device was manufactured in 2008 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed item.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII ARTICULAR INSERTER/EXTRACT
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11391233
MDR Text Key233951211
Report Number1020279-2021-01715
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010214393
UDI-Public03596010214393
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440194
Device Catalogue Number71440194
Device Lot Number08CM05845
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-