Model Number N/A |
Device Problems
Loose or Intermittent Connection (1371); Failure to Cut (2587); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
|
|
Event Description
|
It was reported that during operation of the device the width plate was becoming loose.There was impact to the graft that was harvested and it was not able to be used.It is unknown at this time if additional graft was required as no additional information is available.No other adverse event was reported as a result of the event.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the width plate screw was replaced and resolved the reported issue.Device is used for treatment.A definitive root cause cannot be determined.The event cannot be confirmed.
|
|
Event Description
|
There is no additional information.
|
|