Catalog Number V041441000000 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
As a result of this malfunction, the potential for surgical intervention exists to preclude injury or illness that would necessitate medical or surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event a customer reported that an eddy irrigation tip broke during use.No injury reported.
|
|
Manufacturer Narrative
|
Two tips received in two pieces to share between three prs (b)(4).Tip 1: broken at 25 mm point of passiv part and at 22 mm point of active part.Part in between is missing.No smooth breakage; melted; breakage due to thermal influences.Tip cavi 3.Tip 2: broken approx.At 28 mm point of passive part; active part measuring approx.27,80mm no further broken parts received.Tip cavi 1.Measured with measuring gauge mitutoyo cd-15 cpx valid till 06/2021; no smooth breakages; melted; breakages due to thermal influences.Misuse.
|
|