• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 14 DAY; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71940-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abscess (1690); Burn(s) (1757); Skin Infection (4544)
Event Date 10/07/2017
Event Type  Injury  
Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.The reported complaint is related to skin irritation.Dose audit and environmental monitoring reports were reviewed for issues relating to sterility of the product.The sensor component has no impact on product sterility and therefore, sensor components dhrs were not reviewed.Sensor kit dhrs have been reviewed to assess the manufacturing process, which includes the application of the adhesive to the puck.A dhr (device history review) for the freestyle libre sensor kit was reviewed and the dhr showed the freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.The dose audit report covers the required time period.A tripped trend review was completed for the reported complaint and freestyle libre sensors and no trips were observed.Complaint data analysis has been reviewed for this product and issue.No adverse trends have been identified.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported experiencing an infection while wearing an adc freestyle libre sensor.Customer further reported that after 7 days of wear, he "felt the skin burning and a rash" and noticed "the skin was burned until the elbow".Customer had a friend remove the sensor after 10 days of wear due to discomfort and upon removal, customer noted "the skin was ripped off" and the skin had "been detaching all over the upper arm".Customer further noted "2nd degree burning, blister with liquid, and infected skin" and an abscess at the insertion site.Customer had contact with a healthcare professional and was provided flammazine (topical antibiotic) for treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 14 DAY
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key11398427
MDR Text Key234231574
Report Number2954323-2021-05590
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date06/30/2018
Device Model Number71940-01
Was Device Available for Evaluation? No
Date Manufacturer Received10/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-