Model Number PM3242 |
Device Problem
Failure to Disconnect (2541)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
The patient was presented in the hospital for a generator change.During the procedure, it was observed the right ventricular (rv) lead could not be removed from the header of the pacemaker.After several efforts, the rv lead was released from the pacemaker.The patient was in stable condition.
|
|
Manufacturer Narrative
|
The reported complaint of rv-lead difficult to remove from header was not confirmed.No v-connector anomalies were found.The dimensions of the v-connector port were in specifications.The v-connector was tested for the force it takes to remove and insert leads.Lead removal and insertion forces were normal in specification.No blood or foreign material was found in the connector port that would cause the lead to become stuck.The setscrew had been untightened and removed indicating the setscrew was not the cause of the problem.No device anomalies were found that could have caused this problem.An over-sized lead is a possible cause.
|
|
Search Alerts/Recalls
|