• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO BIPAP W/HUM/CELL, DOM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO BIPAP W/HUM/CELL, DOM; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX700H11C
Device Problem Increase in Pressure (1491)
Patient Problem Atrial Fibrillation (1729)
Event Date 02/09/2021
Event Type  malfunction  
Event Description
A user with an existing medical history of atrial fibrillation alleged the inspiratory pressure on his bi-level positive airway pressure (bipap) device increased suddenly during use, resulting in an episode of atrial fibrillation and admission to a hospital overnight for observation.The manufacturer received the device for investigation and could not confirm the allegation the device inspiratory pressure increased above the prescribed pressure.There were no operational issues, and the device passed all testing.The manufacturer concludes the device did not malfunction as alleged, and that no further action is necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP W/HUM/CELL, DOM
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key11400253
MDR Text Key234308191
Report Number2518422-2021-00551
Device Sequence Number1
Product Code BZD
UDI-Device Identifier00606959025691
UDI-Public00606959025691
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX700H11C
Device Catalogue NumberDSX700H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2021
Initial Date FDA Received03/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age67 YR
-
-