Model Number 66800041 |
Device Problems
Material Rupture (1546); Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Our reference number: (b)(4).
|
|
Event Description
|
It was reported that during the use of the versajet, a patient was undergoing an excision for 3rd degree burns and disassociation of the hose from the water flow at the lock (sterile part) was observed.The problem occurred after 30 minutes of use.The intervention was completed with a new versajet kit.There were no consequences for the patient.
|
|
Manufacturer Narrative
|
The device, used in treatment, was returned for evaluation.A relationship between the report event and device could be established.A visual inspection was performed and showed the feed line hose was received detached from the pump cartridge.There did not appear to be any residual adhesive within the feed line where it would connect to the pump cartridge fitting.Root cause is determine to be a manufacturing process error.This failure is considered to be isolated.Functional inspection was performed and showed the feed line was connected to the feed line fitting within the pump cartridge.The unit then primed and cut.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
|