ST. JUDE MEDICAL 4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number RFG-NT-2000
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2021
Event Type  malfunction  
Manufacturer Narrative
One neurotherm rf generator was received for repair.The nt generator was connected to an external power source and the generator doesn¿t power on successfully.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances were identified that are related to the reported event.Based on the provided information and the observed symptoms, the issue was isolated a burnt capacitor (c49) observed on the rf control board.
 
Event Description
This report is to advise of an event observed during analysis confirming burnt components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4 LESION NT2000 PAIN MANAGEMENT RF GENERATOR
Type of Device
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key11411252
MDR Text Key234634363
Report Number2184149-2021-00064
Device Sequence Number1
Product Code GXD
UDI-Device Identifier05415067022417
UDI-Public05415067022417
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K111576
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRFG-NT-2000
Device Catalogue NumberRFG-NT-2000
Device Lot Number2541-13
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-