• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY MAILLEFER X-SMART PLUS CONTRA ANGLE; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY MAILLEFER X-SMART PLUS CONTRA ANGLE; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number A103200000000
Device Problems Detachment of Device or Device Component (2907); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a x-smart plus contra angle would not hold files.No injury resulted.
 
Manufacturer Narrative
X-smart plus cartridge: bad maintenance (user), there is some rust on the cartridge.X-smart plus wave washer: various mechanical problem, rusted.X-smart plus bearing retainer: various mechanical problem, rusted.X-smart plus head cap: various mechanical problem, rusty.X-smart plus drive shaft: various mechanical problem, rotation not fluid.X-smart plus head assembly other rust.Replaced 1 x xp head cap h1033c1051050, 1 x xp bearing retainer h1033c1051113, 1 x xp cartridge h1033c1051015, 1 x xp wave washer h1033c427a360, 1 x xp drive shaft h1033c1051022 and 1 x xp head assembly h1033c1051011.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART PLUS CONTRA ANGLE
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ  1338
MDR Report Key11418671
MDR Text Key243702444
Report Number8031010-2021-00028
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup
Report Date 10/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA103200000000
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-