The reported event was unconfirmed because the reported failure could not be reproduced.The device met relevant specifications.Visual evaluation of the returned sample noted 6 unopened, unused silicone foley with manufacturing lot number ngeu0860.Visual inspection of the sample noted no obvious visible defects.All of the lumens were noted to be lined up properly within the catheters and no blockages were obvious when observed visually.The drainage lumen was flushed with methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) for all the catheters and all catheters drained normally with no obstruction.No root cause could be found because the reported event was unconfirmed.The device history record review was not required as the reported event was unconfirmed.A labeling review is not required as the reported event is unconfirmed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|