• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number ROTAFLOW
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2021
Event Type  malfunction  
Manufacturer Narrative
The initial failure description was that the rotaflow "display went black during patient treatment.The device was restarted and worked normally again.A getinge service technician was onsite on (b)(6) 2021.To test the affected rotaflow.The technician could not reproduce the reported failure after several start-ups.The rotaflow is working as intended.The rotaflow risk analysis version v06 chapter h1.1.1.34; h1.1.1.37 (dms# 2020689 was reviewed on (b)(6( 2021, with following most probable root cause: h1.1.1.37: malfunction of the microcomputer system: internal cpu failure non-volatile memory failure volatile memory failure in safety relevant variables watchdog failure time base failure on other device parts (rotary knob, display, etc) h1.1.1.34: display defects: wrong cleaning a device history review (dhr) was performed and the dhr does not show any abnormality or issue that is related or can have led to the customer complaint.Based on these investigation results the reported failure could not be confirmed.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.
 
Event Description
It was reported that during patient treatment the display went blank.The customer restarted the rotaflow.After the restart, the device was working as intended again.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key11441560
MDR Text Key245318360
Report Number8010762-2021-00156
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW
Device Catalogue Number701046405
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/12/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-