During an implant procedure, both a high pacing impedance and a high defibrillation impedance were observed on the device.The right ventricular (rv) lead was attempted to be disconnected from the device, but had difficulties being removed from the header.The rv lead was able to be successfully disconnected and then was attempted to be re-inserted into the header, but was unable to be re-connected.The device was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
|
The reported field events of a set screw connection issue and high pacing lead impedance (pli) were verified, however, these events were consistent with the procedure.The right ventricular (rv) df4 set screw was revealed to be completely backed out from the set screw thread, which is a result of unscrewing the set screw too far.When the screw was re-engaged with the connector block, the screw operated normally in affixing the test lead to the header.It is believed that the high pli observed in the field was likely due to the lead connection issue.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Telemetry, pacing, sensing, impedance, high voltage (hv) output, hv shock, and patient notifier were all tested on the bench and no anomalies were detected.
|