• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. VERSAJET EXACT ASSY, 45 DEGREE X 14MM; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. VERSAJET EXACT ASSY, 45 DEGREE X 14MM; LAVAGE, JET Back to Search Results
Model Number 66800041
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2021
Event Type  malfunction  
Event Description
It was reported that, versajet ii handpiece was totally disconnected from the tubing and there was no way to reassemble, it was leaking everywhere.No case involved; therefore, no patient involvement.
 
Manufacturer Narrative
The device, was not used in a procedure was not returned for evaluation with all additional information provided we have not been able to establish a relationship between the reported event or determine a root cause.The manufacturing records show no evidence that the product did not meet the specification at the time of manufacture.A complaint history review found other related failures.Probable cause may be a component failure or connection issue.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET EXACT ASSY, 45 DEGREE X 14MM
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11450019
MDR Text Key238814179
Report Number8043484-2021-00525
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00040565124681
UDI-Public00040565124681
Combination Product (y/n)N
PMA/PMN Number
K110958
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 03/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/11/2022
Device Model Number66800041
Device Catalogue Number66800041
Device Lot Number50895161
Was Device Available for Evaluation? No
Date Manufacturer Received03/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-