Model Number 16-02-80 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).The customer requested deep disinfection in order to solve the reported contamination.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera.The laboratory report confirming the reported contamination has been provided to livanova.There is no known patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: through follow up communication under previous contamination cases from the same hospital, livanova learned that the device was cleaned regularly per the instruction for use and that it was placed outside the operation theater.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|