Model Number 3186 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/26/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that during the patient's permanent scs system implant surgery, the physician was unable to implant the lead due to the white cap of the stylet falling off making it impossible to use the stylet.The surgery was abandoned.
|
|
Manufacturer Narrative
|
An abandoned procedure was reported to abbott.The white cap of a lead fell off and the lead was not steerable during procedure.The results of the investigation are inconclusive as the product was not returned for analysis.As a result of this issue, a supplier corrective action report has been requested.
|
|