• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL COMPIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL COMPIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMC2QQ
Device Problems Premature Discharge of Battery (1057); Failure to Charge (1085); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was hospitalized due to several ventricular tachycardia (vt) and ventricular fibrillation (vf) episodes that were appropriately treated with over 100 shocks by the cardiac resynchronization therapy defibrillator (crt-d).It was further noted that the crt-d battery had reached end of service (eos) prematurely during the vt/vf therapy.Alerts for elevated charge time and charge circuit timeout were triggered due to the high number of shocks.The crt-d was explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated an alert for excessive charge time eos (end of service).Analysis of the device memory indicated charge circuit timeout.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPIA MRI QUAD CRT-D SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
MDR Report Key11474263
MDR Text Key239455009
Report Number9614453-2021-01002
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/14/2020
Device Model NumberDTMC2QQ
Device Catalogue NumberDTMC2QQ
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/04/2021
Initial Date FDA Received03/12/2021
Supplement Dates Manufacturer Received03/14/2021
Supplement Dates FDA Received03/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5076-52 LEAD, 6935M62 LEAD, 439878 LEAD; 5076-52 LEAD, 6935M62 LEAD, 439878 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
-
-