• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX54OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE MTL INS NEUT36IDX54OD; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS Back to Search Results
Model Number 1218-87-354
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Arrhythmia (1721); Foreign Body Reaction (1868); Headache (1880); Pyrosis/Heartburn (1883); Itching Sensation (1943); Pain (1994); Rash (2033); Synovitis (2094); Tinnitus (2103); Discomfort (2330); Ambulation Difficulties (2544); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 08/17/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the patient that " i had my right hip & knee replaced (b)(6) 2006 and my left hip replaced (b)(6) 2007.I began having left hip problems in 2016.I went to the original operating doctor to have it checked.At that time he did not find a solution for my discomfort.He did, however take a cobalt test at that time.It came back at 7.2.This year he took another cobalt test on (b)(6) 2020 and the results were 16.4.After review of medical record patient was revised to address metallosis.Revision notes stated that an inflammatory fluid in the joint and staining of the synovium and synovitis was encountered.Burnishing on the femoral components was noted.Head and liner were removed.Added patient's medical history, laboratory data, dor, revision hospital, and patient's dob.Corrected patient's initials and surgeon.Doi: (b)(6) 2006 - dor: (b)(6) 2020 (right hip).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE MTL INS NEUT36IDX54OD
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key11474700
MDR Text Key243122971
Report Number1818910-2021-04997
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K000306
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2009
Device Model Number1218-87-354
Device Catalogue Number121887354
Device Lot Number1235914
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
APEX HOLE ELIM POSITIVE STOP.; ARTICULEZE M HEAD 36MM +1.5.; PINNACLE 300 ACET CUP 54MM.; SUMMIT POR TAPER SZ5 HI OFF.
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-