Model Number 8220 |
Device Problems
Break (1069); Contamination (1120); Crack (1135); Device Markings/Labelling Problem (2911); Mechanics Altered (2984); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The affected device has been received and an evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
It was reported that the device has a label that is peeling.No additional information was provided.There was no patient involvement.
|
|
Manufacturer Narrative
|
The investigation is still pending.A follow up report will be submitted once the failure investigation has been completed.
|
|
Event Description
|
It was reported that the device has a label that is peeling.No additional information was provided.There was no patient involvement.
|
|
Manufacturer Narrative
|
A review of the complaint history record was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.A review of the device history record showed the device had a manufacture date of 27jul2012.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.
|
|
Event Description
|
It was reported that the device has a label that is peeling.No additional information was provided.There was no patient involvement.
|
|
Event Description
|
It was reported that the device has a label that is peeling.No additional information was provided.There was no patient involvement.
|
|
Manufacturer Narrative
|
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that they replaced lifted keypad.Replaced broken rear case,front case and right iui.Replaced contaminated patient connector pcb and left iui.The failure code othe was used to track the alaris software version as received from the customer and the software version when device was sent back to the customer.It does not reflect a device failure or represent any risk to the patient.A review of the device history record showed the device had a manufacture date of 27jul2012.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.Based on the findings, service determined that the probable cause of the reported issue was due to damage of the keypad 8210/8220.A review of the complaint history record in trackwise and sap was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.
|
|
Search Alerts/Recalls
|