PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR Back to Search Results
Model Number 861290
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2021
Event Type  malfunction  
Event Description
It was reported to philips that the device was not recognizing the pads were attached.The alleged failure was observed while the device was in use on a patient, however, no adverse patient impact was reported by the customer.
 
Event Description
It was reported to philips that the device was not recognizing the pads were attached.The device was received by bench repair.An evaluation was performed by an authorized bench technician and the reported issue was confirmed and traced to a faulty therapy port and processor pca.Upon conclusion of the evaluation, it was determined that this was a malfunction of the therapy port and processor pca.The therapy port and processor pca were replaced to resolve the reported issue and the device was returned to the customer site.No further evaluation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
elizabeth proctor
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key11477584
MDR Text Key239750934
Report Number3030677-2021-10214
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date02/17/2021
Initial Date Manufacturer Received 02/17/2021
Initial Date FDA Received03/12/2021
Supplement Dates Manufacturer Received02/26/2021
Supplement Dates FDA Received11/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-