• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number D2C4063K
Device Problems Crack (1135); Material Discolored (1170)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This event occurred during the unspecified date and month of year 2021.The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that ¿hairline cracking¿ was observed on the fill ports of two (2) folfusors lv (large volume).This issue was identified prior to use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: two (2) actual devices were received for evaluation containing fluid in the bladder.Visual inspection was performed which observed vertical cracks located on the fill port.The reported condition was verified for both samples.The cause of the condition could not be determined, however, the probable cause of the cracks may likely be due to the use of excess alcohol solution to clean the fill port either prior, during or after the filling step.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11477645
MDR Text Key239735115
Report Number1416980-2021-01301
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412474335
UDI-Public(01)00085412474335
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 04/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberD2C4063K
Device Lot Number19K034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2021
Initial Date Manufacturer Received 02/16/2021
Initial Date FDA Received03/12/2021
Supplement Dates Manufacturer Received04/05/2021
Supplement Dates FDA Received04/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-