Model Number 26926
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2021
Event Type  Injury  
Event Description
It was reported that the stent was found to be shorter than what was labeled on the packaging, and additional intervention was required.An 8mm x 150mm x 130cm innova self-expanding stent was selected for use in a stenting procedure.During the procedure, the stent was implanted, but it was found to be shorter than what was labeled on the packaging.The implanted stent length was measured to be 110mm, which was 40mm shorter than expected.An additional 8mm x 60mm x 130mm stent was implanted with 10mm of overlap in order to cover the missed segment.The procedure was completed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11478975
MDR Text Key239682534
Report Number2134265-2021-03077
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08714729874225
UDI-Public08714729874225
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26926
Device Catalogue Number26926
Device Lot Number0026289708
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2021
Initial Date FDA Received03/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/02/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-