• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY ART ISRT ASSM TOOL; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY ART ISRT ASSM TOOL; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74018911
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: case (b)(4).
 
Event Description
It was reported that, during a tka when trying to put poly in, the tabs on the poly gun did not align up with the dimples in the poly so it would not insert the journey art isrt assm tool the poly.The procedure was completed without delay using a smith-nephew back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment was returned for evaluation: a visual inspection was conducted and confirmed that the journey art isrt assm tool the tabs on the poly gun were not aligning correctly.The instrument shows wear and tear which could have caused it to not function correctly.The manufactured date for this device is 2017.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.Case-2021-00042375-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY ART ISRT ASSM TOOL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11480140
MDR Text Key239666705
Report Number1020279-2021-02161
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010557292
UDI-Public03596010557292
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018911
Device Catalogue Number74018911
Device Lot Number17EPG0017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Initial Date Manufacturer Received 02/18/2021
Initial Date FDA Received03/15/2021
Supplement Dates Manufacturer Received04/05/2021
Supplement Dates FDA Received04/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-