As reported by the biomedical engineer, a v044 entry was logged at the time of event.This entry logged indicates a pressure drop at the rolling diaphragm.Unfortunately, as neither the logfile nor any further information such as a service report was made available for investigation, the reported symptom could not be reproduced, and the exact root cause could not be determined.An auxiliary vacuum pressure is needed to operate the valves that control the ventilation cycles and to keep the ventilator diaphragm in place during piston movement.The system reacts with a safety shutdown of automatic ventilation and a corresponding alarm upon insufficient vacuum pressure.A drop in the vacuum pressure may be caused by a number of conditions and some of them are not related to a device malfunction of persisting nature.For example, if the control tube of the peep valve comes off during interaction with the device, will be recognized by the user and re-attached, this leaves no traces for a following device check.Also in the specific case it was reported that the self-test in follow-up to the event passed without any indications for a device malfunction.Dräger finally concludes that the system responded as intended upon a deviation in the auxiliary functions - ventilation was shut down to prevent from damages to the system and, the user was alerted to this condition by means of a corresponding alarm.Manual ventilation remains available.
|