MEDTRONIC ADVANCED ENERGY (SALIENT) PLASMABLADE X; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number PS210-030S-LIGHT
Device Problems No Device Output (1435); Failure to Shut Off (2939)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2021
Event Type  malfunction  
Manufacturer Narrative
Device was returned and is pending analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a manufacture representative (rep) regarding a generator and a handpiece.It was reported that the cut and coagulation button were continuing to run without being pressed.They swapped generator and it was an out of box failure.There was a delay of less than an hour.There was no impact to patient outcome.
 
Manufacturer Narrative
H3: analysis summary: investigation was concluded there were no failures observed upon testing the device for its functionality and worked as expected.Mohammad also mentioned the solder flux could have been remotely the reason of this issue but the device still worked.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMABLADE X
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MEDTRONIC ADVANCED ENERGY (SALIENT)
180 international drive
portsmouth NH 03801
MDR Report Key11486224
MDR Text Key240118017
Report Number1226420-2021-00061
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00763000279172
UDI-Public00763000279172
Combination Product (y/n)N
PMA/PMN Number
K193529
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model NumberPS210-030S-LIGHT
Device Catalogue NumberPS210-030S-LIGHT
Device Lot Number2005044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Date Manufacturer Received04/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-